Ağırlık:
25 kg
Akış Hızı:
100 - 300 mL/min
Alarm :
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analog çıkışlar:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Boyut (Y x G x D):
690 mm x 465 mm x 330 mm
Cihaz havası:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Demineralize su:
For rinsing / dilution
Dijital çıkışlar:
Optional: Modbus (TCP/IP, RS485)
Doğrulama:
Automatic; frequency freely programmable
Doğruluk:
Better than 2% full scale range for standard test solutions
Drenaj:
Atmospheric pressure, vented, min. Ø 32 mm
Enerji:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Garanti:
24 aylar
Girişimler:
Metal ions like Aluminium Al(III), Bismuth Bi(II), Cadmium Cd(II), Chromium Cr(III), Cobalt Co(II), Copper Cu(II), Iron Fe(II), Iron Fe(III), Lead Pb(II), Manganese Mn(II), Magnesium Mg(II), Mercury Hg(II), Palladium Pd(II), Platinum Pt(II), Silver Ag(I), Tin Sn(II), Zinc Zn(II). Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Güç kaynağı:
100 - 240 VAC, 50/60 Hz
Kalibrasyon:
Automatic, 2-point; frequency freely programmable
Koruma sınıfı:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Kutudakiler:
EZ1027 Nickel Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and Empty Containers for Reagents with Fittings
Malzeme:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Model:
EZ1027
Numune akışı sayısı:
1, 2, 4, or 8
Numune basıncı:
By external overflow vessel
Numune kalitesi:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Numune sıcaklığı:
10 - 30 °C
Ölçüm aralığı:
10 - 500 µg/L Ni
10 - 500 µg/L Ni
Optional:
5 - 125 µg/L
5 - 250 µg/L
80 - 2000 µg/L (with internal dilution)
160 - 4000 µg/L (with internal dilution)
1 - 50 mg/L (with internal dilution)
Ölçüm metodu:
Colorimetric measurement using DMG colour solution
Ölçüm süresi:
10 minutes (dilution + 5 min.)
Ortam sıcaklığı:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Otomatik temizleme:
Evet
Parametre:
Nickel Ni(II), dissolved
Reaktif gereksinimi:
Keep between 10 - 30 °C
Saptama limiti:
≤ 5 µg/L
Sertifikalar:
CE compliant / UL certified
Toprak bağlantısı:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²